Power supply : 220V ± 10% V 50Hz
Dimensions : Approximately 510mm × 950mm × 1310mm (width × depth × height)
Power : Approximately 3.5KW
Refrigerant : R404a
Wind speed : 0.3m/s to 0.5m/s (adjustable)
FZ/T 73036 73054 Textile Moisture Absorption Heat Generation Performance Tester
Introduction
It is used for testing the moisture absorption and heat generation performance of textiles, and can also be used for other temperature inspection tests.
Applicable standards
The requirements of standards such as GB/T 29866-2013, FZ/T 73036-2010, FZ/T 73054-2015 and Appendix A.
Product Features
1. Use high-precision temperature sensors;
2. Four-channel detection, supporting simultaneous testing of three samples;
3. Generate temperature variation curves during the test, and automatically calculate the results to produce a report.
Control system composition
1. Temperature measurement: Pt100 platinum resistance;
2. Control Device: Programmable temperature and humidity controller TEMI580. It can display the set parameters, time, heating element and other working status, and also has the functions of automatic test operation and PID parameter self-tuning. Just setting the temperature can realize the automatic operation function of the refrigeration machine. The control system uses an intelligent control software system, which can automatically combine the working conditions of refrigeration, heating, etc. subsystems, thereby ensuring high-precision control within the entire temperature and humidity range and achieving the purpose of energy saving and consumption reduction. The complete detection device can automatically perform detailed fault display, alarm, such as when the test chamber has an abnormality, the controller automatically displays the fault status.
3. Screen display: Set temperature; Measured temperature; Heating status, time, temperature curve, and various alarm indications.
4. Accuracy setting: Temperature: 0.1℃
5. Program capacity: 100 items, total program segments: 1000 segments, maximum single-step program duration: 99 hours and 59 minutes; Programs can be looped and linked between each other.
6. Operating mode: Constant operation, program running;
Safety protection device
1. Studio over-temperature protection;
2. Short-circuit protection for the heater;
3. Fan overload protection;
4. Compressor overpressure protection;
5. Compressor overload protection;
6. Leakage protection;
7. Safe and reliable grounding device;
Technical Specification
Temperature increase value test range and accuracy |
0 - 100℃, resolution 0.01℃ |
Test range and accuracy for average temperature increase value |
0 - 100℃, resolution 0.01℃ |
Studio dimensions |
350mm × 300mm × 400mm (width × depth × height) |
Detection and testing functions |
Utilizing four-channel detection, temperature range 0 - 100℃, resolution 0.01℃; supports simultaneous testing of three samples; generates a temperature variation curve after test completion, with automatic result calculation and report generation |
Temperature control range |
0 - 100℃, resolution: 0.01℃ |
Temperature fluctuation |
≤ ±0.5℃ |
Relative humidity control |
30% to 90% ± 3% |
Wind speed |
0.3m/s to 0.5m/s (adjustable) |
Test time control |
0 minutes: 1 second to 99 minutes: 59 seconds; resolution: 1 second; test error: ±1 second |
Cable threading hole |
One on the side of the test box, diameter 50mm |
Hollow glass observation window |
Dimensions approximately: 200 × 250mm |
Door |
Single door; door sealing with double-layer silicone rubber strips |
Condensate water drain outlet |
Equipped on the box body |
Test chamber interior and box body material |
Test chamber interior: 1mm thick SUS304 stainless steel plate; box body: 1mm thick stainless steel |
Air duct interlayer components |
In the air duct interlayer at one end of the studio, heating/humidification devices, refrigeration evaporators, blower motors, fan blades and other components are installed |
Insulation material |
Polyurethane foam with a thickness of 100mm; excellent insulation effect, no frost or condensation on the outer surface of the test chamber |
Control and protection system |
Continuous PID control; uses SSR solid-state relays as heating actuators (safe and reliable); with a separate over-temperature protection system |
Compressor |
The core component of the refrigeration system is a fully enclosed compressor from France (Taikang), forming a refrigeration system to meet the cooling requirements of the working chamber. The refrigeration system consists of a high-pressure refrigeration cycle and a low-pressure refrigeration cycle, with the connecting container being the evaporator. The evaporator of the low-pressure cycle functions as the condenser of the high-pressure cycle |
Oil separator |
Equipped with the "Aico" ALCO oil separator from Europe and America (to ensure the compressor has sufficient refrigerating oil and avoid performance reduction of heat exchangers) |
Refrigeration evaporator |
Located in the air duct compartment at one end of the test chamber; ventilated by a blower motor for rapid heat exchange |
Energy regulation measures |
While ensuring main technical indicators, the system's refrigeration capacity is adjusted according to different cooling rates and temperature ranges. Adopts additional energy regulation measures such as evaporator temperature regulation, energy regulation, and hot air bypass energy regulation to reduce equipment energy consumption while meeting main technical indicators |
Air duct system |
Equipped with an internal circulation air supply system for high uniformity. Heaters, refrigeration evaporators, fan blades and other devices are distributed in the air duct compartment at one end of the chamber. A fan circulates internal air: air in the chamber is sucked into the air duct from the bottom, heated and refrigerated, then blown out from the top of the air duct. Air that has undergone heat exchange with samples is sucked back into the air duct for repeated circulation to achieve temperature setting requirements |
Refrigerant |
R404a |
Power |
Approximately 3.5KW |
Dimensions |
Approximately 510mm × 950mm × 1310mm (width × depth × height) |
Power supply |
220V ± 10% V 50Hz |
![]() |